Mark Allen wrote:
You're right Al - great test questions.
Let me add a bit more -
On the low oxygen concentration it's a bit more complex. Each train has to have a concentration monitor. If oxygen falls to <91%, the train in question must be automatically closed off from the system to prevent contamination downstream. The Supply Source Isolating Valve is BOTH manual and Automatic for this reason. That is unique to NFPA - it was not part of the ISO document.
Your reading of the alarms is also correct - there are a LOT of them. Because each train is equipped with a manual/automatic valve, the valve itself must be provided with an signal (5.1.3.5.11.12 (5)). If you close the valve manually, it will give an alarm e.g. "Oxygen Concentrator Supply Source Isolated" signal. This is a Master alarm signal only - unusually it is not Local and Master, which is because Locally you will be able to physically look at the valve. If oxygen falls below 91%, the alarm e.g. "Low Oxygen Concentration, Oxygen Concentrator Supply Source X" will be activated at Local and Master panels AND because the Supply Source valve will close automatically, the "Oxygen Concentrator Supply Source Isolated" alarm at the Master will ALSO be activated.
I've not updated my "Oxygen On Site" booklet yet, but I'll try to make the alarms clear when I do. It's not a classic sequence like we are used to.
All the comments about this subject have been grate
I want your opinion about the vale requirement that has to be both manual/automatic, at the end of last year I was verifying a O2 producer and it did not have one vale with the two functions, it has one manual valve and one automatic valve, the automatic was able to activated the local and master alarm signal but the manual did not, how ever this system had a flowmeter unit that also was able to activate the automatic valve and the signals, so if any one close the manual valve the flowmeter unit was able to sense that the generator was isolated and in a sense it could notice that the manual valve was manually close and that the oxygen concentrator was isolated and activated the master alarm signal.
Could you think that this arrangement accomplish the goal of the both manual and automatic valve and signals?