So they give some caveats in this manual regarding when the calibration interval can be moved out to every 6 months.
The calibration interval can be extended to 6 months (maximum), when these requirements can be met:
1. The monitors are in a controlled environment
2. The monitors are in good working condition
3. The CO sensor is replaced every two (2) years
The Beacon Medaes manual also listed a few instances in which calibration must be performed at more frequent than 6 month intervals:
* Recalibration of the CO transmitter is required prior to 6 months if any of the following conditions exist:
• Last Calibration date cannot be verified or is unknown
• Sensor Overexposed to target or interfering gases • Instrument has been subjected to misue or abuse
• Sensor have been newly replaced or instrument has received any type of servicing
• Whenever sensor response is in question
So if you are a hospital facility with the ABL-4021 monitor it would seem you can push calibrations out to every 6 months provided you replace the sensor every 2 years, and meet the conditions in the Beacon Medaes manual (provided it is a Beacon Medaes med air system). At least that is my takeaway