5.1.3.8* Waste Anesthetic Gas Disposal (WAGD).
5.1.3.8.1* Sources. WAGD sources shall be chosen in consultation with the medical staff having knowledge of the requirements to determine the type of system, number and placement of terminals, and other required safety and operating devices.
5.1.3.8.1.1 WAGD shall be permitted to be produced through the medical–surgical vacuum source, by a dedicated producer, or by venturi.
5.1.3.8.1.7 If WAGD is produced by a venturi, the following shall apply:
(1) The venturi shall not be user-adjustable (i.e., require the use of special tools).
(2) The venturi shall be driven using water, inert gas, instrument air, or other dedicated air source.
(3) Medical air shall not be used to power the venture.