Updating Existing Area Alarm Panel

  • Thursday, October 12, 2017 2:44 PM
    Message # 5311101
    Al Moon (Administrator)

    You have a project to add medical air to an existing area alarm panel that has oxygen and vacuum . This existing area alarm panel does not meet the 2012 edition of the 99 code. See below:

     

    (Section # 5.1.9.3.5  Area alarm panels for medical gas systems shall provide visual and audible indication in the event a mismatch occurs between the transducer(s) and its associated circuit board(s).)

     

    If the upgrade kit for the medical air alarm is same brand, type, vintage, style and from the same manufacturer as the existing alarm and works the same way as the existing area alarm panel.

     

    Can it pass the requirement of 2012 edition of the 99 code?

     

     

     

    Last modified: Sunday, October 15, 2017 10:26 AM | Anonymous
  • Friday, October 13, 2017 1:22 AM
    Reply # 5311916 on 5311101

     Good question my initial thinking is no you have to upgrade the entire alarm. 

  • Friday, October 13, 2017 6:38 AM
    Reply # 5312188 on 5311101

    I agree with Don.  I think you would :

    (a): have to update the entire panel, or

    (b): have to leave that panel undisturbed and install a separate panel compliant with the NFPA 2018 just for the air (which is an ugly thing to contemplate).

     

    The "A" answer is (a). 

  • Friday, October 13, 2017 7:31 AM
    Reply # 5312210 on 5311101

    Based on our meeting last week I would say replacement is probably in the cards and the right move, that is to keep JC happy. 