I understand both your comments and though the same.
But after reading the section 4 more times ( mostly part #2 ).
I would debate that per part #2.
This 100 bed with all Oxygen DISS Outlets at 53 psig - can not have a non-standard Oxygen Outlet at 83 psig with a DISS connection.
5.1.5.15 Station outlets in systems having nonstandard operating pressures shall meet the following additional requirements:
(1) They shall be gas-specific.
(2) They shall be pressure-specific where a single gas is piped at more than one operating pressure [e.g., a station outlet for oxygen at 550 kPa (80 psi) shall not accept an adapter for oxygen at 345 kPa (50 psi)].
(3) If operated at a pressure in excess of 550 kPa (80 psi), they shall be either D.I.S.S. connectors or comply with 5.1.5.15(4).
(4) If operated at a gauge pressure between 1380 kPa and 2070 kPa (200 psi and 300 psi), the station outlet shall be designed so as to prevent the removal of the adapter until the pressure has been relieved to prevent the adapter injuring the user or others when removed from the outlet.
NOW AS FOR DON'S QUESTION:
CONCERNING A MEDICAL AIR DISS VS DENTAL AIR DISS.
I WOULD ALSO ASK THE NEXT QUESTION.
WHAT DOES AN ASSE 6030 MEDICAL GAS VERSIFIER, DO WHEN HE SEES:
A DISS IA OUTLET ( INSTRUMENTATION AIR ) IN A CLEAN SCOPE AND THIS OUTLET IS NOT SUPPLIED BY N.F.P.A. 99-2002 EDITION - IA ( INSTRUMENTATION AIR ) COMPLIANT SOURCE SYSTEM?