ZVB Location

  • Friday, May 02, 2025 10:06 AM
    Message # 13494533

    Is this an acceptable location for the Zone Valve Box with only a door separating it from the outlets.  

    1 file
  • Friday, May 02, 2025 10:53 AM
    Reply # 13494556 on 13494533

    Yes, this appears to meet all of the zone valve placement restrictions listed in the code.

  • Saturday, May 03, 2025 7:39 AM
    Reply # 13494854 on 13494533

    Chris - would love to hear the background to this question.  It seems entirely straightforward so there must be a backstory.  Reason I am particularly curious is :


    1. A report that I have heard of an accreditation person who argues "a door is not a wall", therefore your valve location as shown would NOT be acceptable.


    2. A proposal that we’ve made to NFPA 2027 (PI 170, to 5.1.4.6) which should clarify this and of which your example is an argument "For" .