Future use valves - for temporary supply service / back feeding

  • Monday, April 26, 2021 8:42 AM
    Message # 10359066
    Cary Darden (Administrator)

    5.1.4.8.3

    Valves for future connection intended for use only as a temporary service feed (back feed) location shall be permitted to be installed with a female adapter and plug.  These valves must be normally closed and only open when a temporary supply is being connected to the system.


    This would allow facilities to install a temporary service feed valve with a threaded connection and plug on the future use side of the valve to facilitate easily providing temporary service to a given area.  These valves would eliminate the need for having to backfeed through a patient room outlet as is typically done.